PEDIATRIC Radiotransparent Electrode, REF/Catalog Number 2603H, Rx ONLY --- COMPARABLE TO PHILIPS MEDICAL HEARTSTART ELECTRODES PLUG STYLE CONNECTOR

FDA Recall Z-0459-2015 · Posted 2014-12-04

Recalling FirmConMed Corporation
Recall StatusTerminated
Product CodeMKJ
Date Initiated2014-11-06
Date Terminated2017-10-30
Product Quantity174,610 (168,090 US and 6,520 OUS) in total.

Reason for Recall

Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury.

Firm Action

ConMed sent an Urgent Device Correction letter dated November 16, 2014, to all affected consignees via UPS Next Day Air. The letter instructed consignees to notify all end users of the correction. For questions or additional information, consignees were instructed to contact ConMed at 1-727-399-5276 or by e-mail at [email protected]. International notices were sent via UPS Priority or equivalent by November 19, 2014.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - USA (nationwide) Puerto Rico, and Internationally to Bosnia, Herzegovina, Brazil, Canada, Great Britain, Israel, Italy, Republic of Korea, Norway, Portugal, Saudi Arabia, and Spain.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.