RadPRO(R) Mobile 40kW Digital X-Ray System - Model: SM-40HF-B-D-C Product Usage: Instrument is used to take diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts on both adult and pediatric subjects. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography
FDA Recall Z-0460-2019 · Posted None
Advertisement
Recalling FirmVirtual Imaging, Inc.
Recall StatusTerminated
Product CodeMQB
Date Initiated2018-11-02
Date Terminated2020-08-19
Product Quantity253 units
Reason for Recall
There is a potential for the RadPRO Mobile 40kW Digital X-Ray System Model SM-40HF-B-D-C to separate from the Tube Head Mount and potentially fall off the system.
Firm Action
The firm initiated the field correction by issuing a field safety notice on 11/03/2018. The letter identified the affected product, problem and action to be taken.
Root Cause
Component change control
Distribution Pattern
Worldwide Distribution - US Nationwide and the countries of, Canada, Italy