Novoste, Beta-Cath 3.5F System, New B-Rail 3.5F Delivery Catheter, with Indicator of Source Train (IST), Novoste Corporate, 3890 Steve Reynolds Blvd, Norcross, GA 30093, Tel: +1770-717-0904, Fax: +1770-717-1283, 1-800-NOVOSTE (1-800-668-6783), www.novoste.com.
FDA Recall Z-0462-03 · Posted 2003-01-30
Advertisement
Recalling FirmNovoste Corporation
Recall StatusTerminated
Product CodeJAJ
Date Initiated2002-01-25
Date Terminated2003-11-17
Product Quantity50 units
Reason for Recall
Mislabeling. Source calibrated incorrectly, resultant dose rate lower than expected.
Firm Action
Customers were notified via Advisory Letter followed up with a Recall Letter and Customer Informational Letter on/about January 24, 2002. Customers were advised to discontinue use of the product and return the affected product to Novoste Corp. to be recalibrated to obtain a correct calibration certificate.
Root Cause
Other
Distribution Pattern
Nationwide. (AR, AZ, CA, FL, GA, IA, ID, IL, KY, MD, MI, MN, MO, NC, NM, NV, NY, OH, PA, TN, TX, VA, WA). No foreign or government accounts involved.