Terminated

Premier brand C. difficile Toxin A, diagnstic test kits, one kit per pouch, one pouch per shipping box.

FDA Recall Z-0463-03 · Posted 2003-01-23

Recall StatusTerminated
Product CodeMCB
Date Initiated2002-12-26
Date Terminated2004-03-29
Product Quantity1,215 kits

Reason for Recall

Due to defective packaging, the firm cannot assure product stability through the labeled expiration date.

Firm Action

The firm mailed notices to their affected customers on 12/26/2002. The consignees were instructed to visually and physically inspect the seals on each microwell pouch for gaps. If properly sealed, the use of the microwells may be continued. Test wells in pouches with defective seals are to be quarantined and destroyed. The customers are asked to report back to the firm. All test microwell plates in defective pouches will be replaced by the firm.

Root Cause

Other

Distribution Pattern

The kits were distributed to end-users and distributors located nationwide; and to foreign distributors located in China, Guatemala, New Zealand, Hong Kong, Mexico, Great Britain, Belgium, Italy, Brazil, Australia, Canada, Thailand, Honduras, and Japan.