CD HORIZON (R) Spinal System, MULTI-AXIAL SCREW FOR 5.5mm ROD W/DUAL LEAD THREAD, Size 6.5mm X 40mm, Mat'l TITANIUM ALLOY, REF X0809439, Rx only, Medtronic Sofamor Danek USA, 1800 Pyramid Place, Memphis, TN 38132 implantation
FDA Recall Z-0464-2012 · Posted 2012-01-11
Advertisement
Recalling FirmMedtronic Sofamor Danek USA Inc
Recall StatusTerminated
Product CodeKWQ
Date Initiated2011-11-30
Date Terminated2012-04-05
Product Quantity252 units
Reason for Recall
One screw may be longer then size etched on the screw.
Firm Action
Medtronic Sofamor Danek USA, Inc sent a Voluntary Field Correction letter dated November 30, 2011, via FedEx to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer.
The firm will visit each consignee to perform a visual check of inventory.
Please utilize the attached questionnare to account for the affected product and to record any patient injuries that may have occurred specifically related to the use of these items.
If you need assistance, or if you have any related questions or concerns please call (800) 876-3133 extension 3173
Root Cause
Mixed-up of materials/components
Distribution Pattern
Nationwide Distribution including the states of AL, PA, IN, CO, MI and IA.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.