Siemens syngo.plaza Radiological Image Processing System Usage: Radiological Image Processing System

FDA Recall Z-0465-2013 · Posted 2012-12-03

Recall StatusTerminated
Product CodeLLZ
Date Initiated2012-06-20
Date Terminated2014-01-14
Product Quantity15

Reason for Recall

Firm has become aware of an unintended behavior when using syngo.plaza and issued an Update Instruction. The Update Instruction SY050/12/S provides a software update and addresses the following issues: - Patient Rename messages for the same patient are being continually processed for several days. This fills up the 'pcvdbsrv' log and potentially affects performance of the system. - Data could

Firm Action

The firm, Siemens, sent "Update Instruction SY050/12/S" on June 20, 2012 to its consignees/customers. The Update Instructions described the product, problem and actions to be taken. It also provides a software update. If you have any questions call 610-219-4834 or email: [email protected].

Root Cause

Software design

Distribution Pattern

Nationwide distribution: USA including states of: CA, FL, GA, MA, MN, MO, NJ, NY, OH, and WI.