Disposable Subdermal Needle Electrode, Corkscrew, Single Use, Sterile; Model Numbers TE/S50715-001-C, TE/S50725-001C, and TE/S50715-001.
FDA Recall Z-0465-2022 · Posted None
Advertisement
Recalling FirmTechnomed Europe
Recall StatusTerminated
Product CodeGXZ
Date Initiated2021-12-08
Date Terminated2024-05-31
Product Quantity117,672 devices
Reason for Recall
There is a high risk of the blue hub detaching from the corkscrew needle, resulting the the needle remaining in the patient's scalp.
Firm Action
The firm distributed an Urgent Medical Device Recall notice dated 12/07/2021 to customers in the United States via email on 12/08/2021. In this notice the firm identifies affected product as being recalled due to the high risk of the needle electrode and lead wire coming loose from the hub, resulting in potential medical intervention to remove the inserted needle electrode, including removal using forceps or incision. The recalling firm requested that customers notify any recipients of recalled product down the distribution chain by forwarding a copy of the notice, that the customer be aware that the firm will schedule a telephone call to initiate the replacement of affected product, and that only distributors return the provided Acknowledgement Form following the scheduled phone call to [email protected]. Customers are asked to ascertain the amount of affected product on hand to be returned with their lot numbers for the scheduled call. Customers with any questions are to contact the recalling firm at [email protected] or by phone at +31 43 608 48 48.
Root Cause
Device Design
Distribution Pattern
US Nationwide distribution in the states of California, Florida, and Ohio.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement