Terminated

1) Accolade II 127 Neck Angle Hip Stem, Size # 3 2) Accolade II 127 Neck Angle Hip Stem, Size #5 Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 USA Telephone # +1 201-831-5000 Stryker France ZAC Satolas Green Pusignan Av de Satolas Green 69881 MEYZIEU Cedex France Intended for cementless application in total or hemiarthroplasty procedures.

FDA Recall Z-0467-2014 · Posted 2013-12-07

Recall StatusTerminated
Product CodeJDI
Date Initiated2013-10-25
Date Terminated2015-03-31
Product Quantity20 units

Reason for Recall

Stryker received a report from the field indicating that an Accolade II Hip Stem Size #3 was identified in packaging with an Accolade II Hip Stem Size #5.

Firm Action

Stryker sent an e-mail to their domestic locations on 10/11/2013. Notification letters and product accountability forms were sent via Fed Ex on 10/25/2013 with return receipt to branches/agencies and hospital risk management, chief of orthopaedics and surgeons.

Root Cause

Packaging process control

Distribution Pattern

Nationwide Distribution including states: AL, GA, IL, and WA.