Terminated

Re-Flex VSP and Low Profile Re-Flex VSP external prosthetic foot modules with integrated shock absorber, Model numbers RFXS0002, RFXS0003, RFLS0006 and RFL50007

FDA Recall Z-0468-2007 · Posted 2007-01-10

Recalling FirmOssur
Recall StatusTerminated
Product CodeISH
Date Initiated2006-11-03
Date Terminated2008-05-28
Product Quantity138 worldwide (86 in US)

Reason for Recall

Inner Tube Breakage: distributed low quality units can lead to breakage of the inner tube, resulting in an abrupt rotation of the shock module, compromising function and stability of the prosthetic foot

Firm Action

By letter on Nov 9, 2006. An initial letter was sent to all involved customers stating the invoice number, the purchase order number and the original shipping date. The problem is explained and the detailed exchange process is outlined. Contact information is included (ref. enclosure 2.1). At the same time as the mailing, all customers are being contacted proactively by a customer service representative.

Root Cause

Other

Distribution Pattern

Nationwide in US plus Worldwide distribution in Canada, Austria, Switzerland, Czech Republic, Germany, Spain, France, United Kingdom, Greece, Israel, Italy, Netherlands and Norway.

Official Source

View FDA Recall →