Hipstar V40 Femoral Stem 127 degrees; N degrees 3 high offset 31 mm;Sterile; Catalog No: 7850-5-006, IMPLANT SIZE 6 , Howmedica,Benoist Girard dt Cie, 203 bd de la Grande Della B.P.8;14291 Herouville Saint Clair

FDA Recall Z-0468-2008 · Posted 2007-12-14

Recall StatusTerminated
Product CodeLPH
Date Initiated2007-09-28
Date Terminated2008-10-27
Product Quantity105

Reason for Recall

Mis-labeling; The warning label "Do not use with heads more than +10mm offset" does not appear on 127 degree stem boxes.

Firm Action

Urgent Product Correction/Recall notification letters were sent by Fed Ex return receipt on 9/28/07, requesting quarantine for Styker representatives to return or re-label.

Root Cause

Packaging process control

Distribution Pattern

Nationwide- USA including states of OR, AZ, NV, MI, AL, MI, FL, MN, and CA