Recall StatusTerminated
Date Initiated2020-10-21
Date Terminated2021-05-05
Product Quantity26 distributed worldwide (4 in US)
Reason for Recall
Some electronic components within the generator were equipped with an incorrect protection diode. As a result, protection against overvoltage may be inadequate, and the system may not function properly in the event an overvoltage situation occurs.
Firm Action
On October 21, 2020, the firm notified customers via Urgent Medical Device Correction letter. The letter informed customers of the hardware error, and that customers could continue to use their system but needed to follow their standard emergency procedures they had in place in the event of system failures. It is recommended to consider reinforcing these emergency procedures until the update has been completed. Siemens advises that an alternative system is available to continue treatment in case of occurrence.
Siemens is implementing a corrective action to all affected users via Update Instruction AX018/20/S. The Siemens Service organization will exchange the affected printed circuit boards of the generator. This modification will bring the voltage back to the center of the tolerance range.
Siemens' service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact the service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented.
Distribution Pattern
Worldwide distribution - US Nationwide distribution including in the states of AR, CA, IA, IL and the countries of China, Germany, India, Indonesia, Italy, Japan, Philippines, Saudi Arabia, South Korea, Taiwan, Turkey, and United Kingdom.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.