Terminated

Sickle-STAT Test Kit, Catalog # SC901, a qualitative test kit for the determination of the presence of hemoglobin S in human blood --- FOR IN VITRO DIAGNOSTIC USE --- FOR PROFESSIONAL USE ONLY --- Contents: 48 pre-filled Sickle-STAT reaction tubes containing sodium hydrosulfite. 1 bottle phosphate buffer containing surfactant and stabilizer, and 1 Instruction leaflet. --- manufactured by: CHEMBIO DIAGNOSTIC SYSTEMS, INC., MEDFORD, NY 11763 --- Accessories available: SC902: Sickle-STAT 5-place Observation rack, SC903: Sickle-STAT Hemoglobin S Control Set; 4 x 1 vials: 2 A/A negative and 2 A/S positive, SC904: Sickle-STAT Phosphate Buffer, 200 ml.

FDA Recall Z-0470-06 · Posted 2006-02-04

Recall StatusTerminated
Product CodeGHM
Date Initiated2005-12-16
Date Terminated2006-06-30
Product QuantityLot SC050305: 347 test kits, Lot SC062805: 410 test kits

Reason for Recall

Firm received one report of a false negative with the positive control. Internal testing by the firm has shown that the kit lot may result in invalid or false negative interpretations.

Firm Action

Chembio phoned the consignees and followed up by faxing the recall letters (dated 12/15/05) & response forms on 12/22/05.

Root Cause

Other

Distribution Pattern

Medical facilities and a few distributors located nationwide, the Dominican Republic, and West Indies (St. Vincent & Grenadine)