Black Mamba Suture Passer
FDA Recall Z-0471-2016 · Posted None
Advertisement
Recalling FirmBiomet, Inc.
Recall StatusTerminated
Product CodeHWQ
Date Initiated2015-12-02
Date Terminated2016-10-27
Product Quantity50
Reason for Recall
There may be a separation
and an insufficient weld between the spring and washer. The weld between the spring and washer may fracture,
causing the Mamba Suture Passer instruments jaw to not actuate properly.
6 complaints have been received in which 3 have been reported with a delay up to 10 minutes.
Firm Action
On 12/2/2015, URGENT MEDICAL DEVICE RECALL NOTICE notifications were sent to the affected distributors and hospital risk managers via courier with instructions to locate and return the affected products. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Root Cause
Nonconforming Material/Component
Distribution Pattern
FL, NJ, WI, UT, TX, KY, KS, OH, GA, LA, AL, NY, NC, AR, MO, VA, CA, MN, IN, HI
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.