Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Hipstar TMZF Cementless Hip Stem Protocol, Product Number: LSP60.

FDA Recall Z-0475-2008 · Posted 2007-12-14

Recall StatusTerminated
Product CodeLPH
Date Initiated2007-09-28
Date Terminated2008-09-05
Product Quantity192 protocols

Reason for Recall

Mis-labeling: An error on page 4 of the Hipstar protocol was found. In the table for compatible Heads, the 40 mm diameter heads with a Neck Length of +12 should have been omitted.

Firm Action

Urgent Product Correction Letters, dated 10/17/2007 were sent on 10/17/07by Federal Express to all Stryker Branches/Agencies who had received the protocol. The branches were instructed to quarantine and destroy the protocols.

Root Cause

Other

Distribution Pattern

Nationwide:The protocol was distributed to 9 Stryker branches/agencies in NJ, MD, IN, MI, MN, IL and TX.

Official Source

View FDA Recall →