Total Toe Instrument Kit, Labeled in part: "Instrument Set, 15mm Toe, Dorsal Flange Combi" Part Number: 9000-3002 The component in question is the hex driver (part number 2009-4001).
FDA Recall Z-0475-2016 · Posted 2015-12-23
Advertisement
Recalling FirmArthrosurface, Inc.
Recall StatusTerminated
Product CodeLZJ
Date Initiated2014-06-09
Date Terminated2016-02-05
Product Quantity3 units
Reason for Recall
An oversight in the implementation of the design change record resulted in outdated components being shipped to customers. The component did not pass a half-cycle sterilization.
Firm Action
The distributors were verbally notified, and replacement Rev F Hex Drivers sent to the locations of the kits. The kits had been moved in preparation for upcoming surgeries, but the Drivers were replaced prior to any additional surgery taking place. No affected product remains in the field. For further questions, please call (508) 520-3003.
Root Cause
Process change control
Distribution Pattern
US Distribution to the states of : PA, WV and CA.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.