Terminated

VITROS Chemistry Products dLDL Reagent-GENs 26 & 27. In Vitro Diagnostic to quantitatively measure LDL Cholesterol (LDLC) Product Code: 6801728 UDI (10758750006793)

FDA Recall Z-0476-2019 · Posted None

Recall StatusTerminated
Product CodeMRR
Date Initiated2018-10-08
Date Terminated2022-10-13
Product Quantity5850 units

Reason for Recall

VITROS Chemistry Products dLDL Reagent GENs 26 and 27, negatively biased results

Firm Action

Ortho Clinical Diagnostics issue the customer notification on October 8, 2018, provides information regarding the potential for biased results when using VITROS dLDL Reagent, Generations (GENs) 26 and 27. Actions to take: For GEN 26: Discontinue using GEN 26 and switch to an alternate GEN of product as soon as possible. Please specify full or partial sales units remaining and indicate credit or replacement on your Confirmation of Receipt form. " For GEN 27: Install ADD DRV 6003 on your VITROS System upon receipt. Following a successful recalibration, resume using VITROS dLDL Reagent, GEN 27 remaining in your inventory. " Complete the Confirmation of Receipt form and return by October 22, 2018. Questions contact our Ortho Care" Technical Solutions Center at 1-800-421-3311

Root Cause

Under Investigation by firm