JuggerLoc/ToggleLoc 2.9mm Drill Guide, Model number 110027357
FDA Recall Z-0477-2025 · Posted 2024-11-15
Advertisement
Recalling FirmBiomet, Inc.
Recall StatusOpen, Classified
Product CodeKDD
Date Initiated2024-10-21
Product Quantity691 (US); 2,045 (OUS)
Reason for Recall
Firm received complaints reporting that guides were warped when pulled out of packaging. This issue may result in extension of surgery to find another product, or may lead to adverse local tissue reaction leading to surgical intervention.
Firm Action
Pending
Root Cause
Process control
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Australia, Brazil, Canada, Singapore, Switzerland.