RayStation standalone software treatment planning system, Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0, 9.1, 9.2, 10.0 - Product Usage: RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments. The system functionality can be configured based on user needs.

FDA Recall Z-0478-2021 · Posted None

Recall StatusTerminated
Product CodeMUJ
Date Initiated2020-10-08
Date Terminated2023-03-30
Product Quantity2768

Reason for Recall

If a region of interest (ROI) or point of interest (POI) that is referenced from an imported plan is missing in the imported RT Structure Set, the reference may become linked to the wrong ROI or POI.

Firm Action

On October 8, 2020, the firm emailed "Field Safety Notice, Medical Device Correction #66704" to affected customers. Customers were made aware of the product issue and asked to take the following actions: " Do not exclude the prescription ROI or POI when exporting from RayStation with the Exclude from export option. " If an RT Structure Set where it is not certain that all referenced bolus and prescription ROIs or POIs are present is imported to RayStation, make sure to check that any intended references to bolus and prescription are correct after import. " Please educate planning staff and all users about this workaround. " Inspect your product and identify all installed units with the above software version number(s). " Confirm you have read and understood this notice by replying to the notification email. This issue will be resolved in the next version of RayStation, scheduled for market release in November 2020 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use. For regulatory information, please contact [email protected]

Root Cause

Software design

Distribution Pattern

US Nationwide distribution.