Terminated

Micro Plate EIA Oral Fluid Cutoff Calibrator which is a component of the Intercept Micro Plate EIA kit.

FDA Recall Z-0479-2009 · Posted 2008-12-23

Recall StatusTerminated
Product CodeLDJ
Date Initiated2008-09-08
Date Terminated2009-05-05
Product Quantity2198 units

Reason for Recall

Results, false-positive test: low % displacement result of 36.0% (spec 39-60%). For further information, please contact the firm at 610-882-1820.

Firm Action

The recalling firm issued a letter entitled "URGENT PRODUCT RECALL NOTICE" dated 9/8/08 to their customers to inform them of the problem and the need to return the product.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution --- USA including states of AZ, CA, CT, DE, FL, GA, ID, KS, KY, MA, MD, MI, NC, NH, NV, NY, PA, VA, and WY, and countries of United Kingdom, Italy, Germany, and Canada.