EMPOWR Partial Knee Peg Drill Guide; Model: 800-06-007.
FDA Recall Z-0479-2021 · Posted None
Advertisement
Recalling FirmEncore Medical, LP
Recall StatusOpen, Classified
Product CodeHSX
Date Initiated2020-06-19
Product Quantity28 units
Reason for Recall
There have been issues experienced with the peg preparation steps, specifically drilling of the Femoral posterior peg. Upon drilling, the shaft portion of the drill may bind within the Posterior Peg drill guide. The extended length of the drill guide and tight tolerances to ensure accurate preparation seem to be contributing to this situation. There is a potential of a delay in surgery if this issue were to recur.
Firm Action
The method of notification was via e-mail to surgical sales agents, and the notices were issued on 06/18/2020. Additionally, a notification to return the affected drills for replacement were sent via e-mail to affected surgical agents on 07/20/20.
The following recommendations should be communicated to the implanting surgeons:
Sales Agents Action:
To avoid the potential binding, the following is recommended:
" Use a plunge mill approach, i.e. plunge the drill in and out of the drill guide bushing a few times slowly until the hole is fully prepared.
" Use of saline or fat to keep the drill lubricated during use.
" Inspect all instrumentation prior to use to identify any wear, breakage or damage.
Along with the notification, the communication includes an acknowledgment form that should be completed and returned to confirm that the instructions provided have been read and understood. THE ACKNOWLEDGEMENT SHOULD BE REVIEWED AND COMPLETED BY JULY 6, 2020.For questions, contact Teffany Hutto by phone at (512) 834-6255 or by email at [email protected].
Root Cause
Other
Distribution Pattern
US: AZ, CA, FL, GA, IL, IN, MI, MN, NY, TX, VA, and WA.. No OUS Consignees.