BEAR Implant (Bridge-Enhanced ACL Restoration); Model/Catalog: 1000;

FDA Recall Z-0479-2026 · Posted 2025-11-07

Recalling FirmMiach Orthopaedics
Recall StatusOpen, Classified
Product CodeQNI
Date Initiated2023-02-02
Product Quantity167 units

Reason for Recall

The field action was initiated in February of 2023 and was limited to the affected lot from sales force trunk stock only.

Firm Action

On January 30, 2023 RE: BEAR Implant Hydration memo emails were sent to sales staff. If you observe an implant that hydrates too quickly (e.g., less than 30 seconds) and appears that it may not remain intact upon insertion into the knee, do not attempt to implant the device. We would recommend you always have a backup device available to address such circumstances. We plan to revise the Tips and Pearls document, ML-1027, to provide further instructions/details on how to properly hydrate the BEAR device. In the meantime, one way to avoid shortened hydration time is to consider hydrating the dense side of the implant first and continue toward the porous side . To date, all reports of this observation have involved Lot # 7008409. As a precaution and to ensure customer satisfaction, we are asking that you return all devices from this lot that are currently in your trunk stock (in your possession). Chris Brown will work with you to coordinate returns as well as replacing these trunk stock devices. Please contact me ([email protected]) or Arjun ([email protected]) if you have questions.

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide distribution.