Animas ezManager Max Diabetes Management System version 2.0.9
FDA Recall Z-0481-2010 · Posted 2009-12-03
Advertisement
Recalling FirmAnimas Corporation
Recall StatusTerminated
Product CodeMRZ
Date Initiated2009-11-02
Date Terminated2010-10-27
Product Quantity20,648 units
Reason for Recall
Printed hard copy of results is not the same as the system screens- carbohydrate to insulin ratio incorrect.
Firm Action
Animas issued an Urgent Medical Device Correction letter dated 11/2/09 to the pump user and another letter to the health care professional that prescribed the pump. The letters explain the issue. The patient letter is accompanied with a CD that contains the upgraded software and instructions to follow.
Root Cause
Software design
Distribution Pattern
Nationwide distribution, and Canada, Cayman Islands, and Denmark.