Atrium Oasis Dry Suction Water Seal Chest Drain, Part Number: 3650-100, DRAIN, OASIS BRU W/AC, 3650 ATS BLOOD RECOVERY

FDA Recall Z-0486-2024 · Posted 2023-12-02

Recall StatusOpen, Classified
Product CodeCAC
Date Initiated2023-09-05

Reason for Recall

Product was re-processed and re-sterilized by a third party using packaging configurations and a sterilization process which are not approved by the manufacture and lack validation.

Firm Action

The recall notification was issued via Fed-Ex and email on October 5, 2023. It requests (1) immediate quarantine and return of all affected product inventory on hand. (2) for any product further distributed, notify customers that have received affected product to remove the affected devices from use and return them. Questions may be addressed to your Getinge representative or call Getinge Customer Support at (603) 880-1433 (press option 1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).

Root Cause

Labeling mix-ups

Distribution Pattern

US: VA