Recall StatusTerminated
Date Initiated2009-10-27
Date Terminated2012-03-22
Product Quantity5046 units (3720 units in US)
Reason for Recall
The recall was initiated because during the manufacturing process of the affected units, the molding process resulted in lock nuts that did not have the level of acceptability required, and therefore, may not withstand the pressures exhibited by the Elements Obturation Unit with which the cartridges are used. More Specifically, there exists a possibility that if the lock nut were to fail, the cart
Firm Action
The recall communications was initiated on November 06, 2009 with the firm forwarding a Recall Letter to the affected consignees via US Postal Service 1st class mail. The recall letter informed the consignees of the reason for recall, the hazard and the affected products. Consignees were instructed to call SybronEndo Customer Care at 1-800-346-3636 directly to handle returns & replacements. In addition, consignees were asked to complete the enclosed Recall Return Form and return it by fax to 909-962-5605.
The recall notice also instructed consignees that are an authorized SybronEndo distributor to identify and recover the affected product lots listed above that may have been shipped to their customers.
Root Cause
Component design/selection
Distribution Pattern
Nationwide and Worldwide to: USA, Canada, Chile, China, Ecuador, Guatemala, Hong Kong, Japan, Malaysia, Singapore, Taiwan and Venezuela
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.