GuardIVa Antimicrobial Hemostatic IV Dressing, REF FP-23-AD008, Manufacturer: Bard Access Systems, Inc. USA. Product Code FP23AD006.
FDA Recall Z-0493-2015 · Posted 2014-12-01
Advertisement
Recalling FirmBard Access Systems
Recall StatusTerminated
Product CodeFRO
Date Initiated2014-11-03
Date Terminated2015-05-12
Product Quantity2532 units
Reason for Recall
Bard Access Systems is recalling GuardIVa (Ref. No. FP23AD006) from lot ASXHT003 because they were non-sterile unit samples which potentially could be used clinically.
Firm Action
A customer letter was sent on 11/3/14 requesting immediate examination of inventory and to quarantine the products for return.
Root Cause
Mixed-up of materials/components
Distribution Pattern
US Distribution to state of: TX, OK, LA, and AR.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement