Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Singles carbides by Meisinger, US-No: 245, Sterile R, cylinder round end; US-No 245; Reference: HMUN245-009; Shank: FG/314; Head Size 1/10mm: 009,Head Length/mm:

FDA Recall Z-0493-2020 · Posted None

Recall StatusOpen, Classified
Product CodeEJL
Date Initiated2018-11-07
Product Quantity2500 burs

Reason for Recall

Incorrect packaging; There is a potential that a 1.0 mm carbide blade was packaged in a 0.9 mm carbide blade packaging.

Firm Action

On 10/29/2018, a medical device removal notice was mailed to the customers instructing them to do the following: 1. check your inventory for burs from the affected batch . Please return any affected instruments. You will receive a swift replacement. 2. If you have further distributed this product, please identify your customers in the response form attached . 3. Please complete the reposne orm and sent it bat to the firm at +1 (303268-5407 or by e-mail at [email protected] 4. For further questions, our customer service is available at +1 (303)268-5407

Root Cause

Employee error

Distribution Pattern

US: NV, PA, FL, SC, IA, NC-