Terminated

Access Progesterone Assay, Part number 33550.

FDA Recall Z-0494-03 · Posted 2003-01-30

Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeJLS
Date Initiated2002-12-06
Date Terminated2003-07-26
Product Quantity6,655

Reason for Recall

Primary tube sampling issue may cause false low results.

Firm Action

Recall letters were sent to customers on 12/6/2002 informing them to not process the sample directly from the primary collection tubes but to pour off a bit into a separate tube. Labs are to incorporate this into their Standard Operating Procedures until the firm can identify another resolution to the problem.

Root Cause

Other

Distribution Pattern

USA and Canada.