Access Progesterone Assay, Part number 33550.
FDA Recall Z-0494-03 · Posted 2003-01-30
Advertisement
Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeJLS
Date Initiated2002-12-06
Date Terminated2003-07-26
Product Quantity6,655
Reason for Recall
Primary tube sampling issue may cause false low results.
Firm Action
Recall letters were sent to customers on 12/6/2002 informing them to not process the sample directly from the primary collection tubes but to pour off a bit into a separate tube. Labs are to incorporate this into their Standard Operating Procedures until the firm can identify another resolution to the problem.
Root Cause
Other
Distribution Pattern
USA and Canada.