15 ML REAGENT CUP, B1037307, SAP 1221037307 for use with ABX PENTRA 400 / PENTRA C400
FDA Recall Z-0494-2018 · Posted None
Advertisement
Recalling FirmHoriba Instruments, Inc dba Horiba Medical
Recall StatusTerminated
Product CodeJJE
Date Initiated2017-11-21
Date Terminated2019-11-04
Product Quantity10 bags of 100 units
Reason for Recall
The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.
Firm Action
The firm initiated their recall by letter on 11/21/2017. The letter stated the following:
"A new production of reagent container compliant for each reference is under process and will be made available very soon. In the meantime and to allow a proper operating of the analyzer ABX PENTRA 400 / PENTRA C400, it Is necessary to push the reagent containers in the reagent racks until they touch the bottom of the rack.
Please share this information with your laboratory staff, and retain this notification as part of your Quality System documentation. It is mandatory for you to complete and return the enclosed response form within 10 days so we
may maintain our records."
Root Cause
Process control
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.