Trach-Eze Closed Suction System Endotracheal Start Kit *** 16F *** REF 220070 *** Catheter Length: 571 mm *** STERILE ***.
FDA Recall Z-0496-03 · Posted 2003-01-30
Advertisement
Recalling FirmSorenson Medical, Inc.
Recall StatusTerminated
Product CodeBSY
Date Initiated2002-11-25
Date Terminated2004-07-12
Product Quantity2790 units
Reason for Recall
Catheters have 2 unvalidated seal product configurations to produce a new hybrid seal that had not been verified or validated.
Firm Action
On 10/25/2002, all consignees were notified by letter.
Root Cause
Other
Distribution Pattern
Distributors in IL, OH, PA. Foreign distribution to Brazil, Czech Republic, Denmark, Guatemala, Italy.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.