WECK AutoEndo5, Automatic Hem-O-Lok Clip Applier, REF AE05ML Automatic Endoscopic Surgical Ligating Clip Appliers are delivery devices for Hem-o-lok ML non-absorbable polymer ligating clips
FDA Recall Z-0496-2020 · Posted None
Advertisement
Recalling FirmTeleflex Medical
Recall StatusOpen, Classified
Product CodeFZP
Date Initiated2019-10-04
Product Quantity34491 devices
Reason for Recall
Certain clip appliers have an increase in misloading and/or jamming related complaints that may be due to a discrepancy in the specification a component. Misloading and/or jamming may lead to clips falling out of the applier, clip breakage, or the jaws becoming locked in a partially closed position (jammed).
Firm Action
The firm notified their consignees by letter on 10/04/2019. The letter explained the issue and requested the user discontinue use and contact the company for instructions for their return. Distributors were directed to notify their customers and collect the affected product for return to Teleflex Medical.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide distribution, including US nationwide (including Puerto Rico).