TrailBlazer Support Catheter, ev3 inc, 4600 Nathan Lane North, Plymouth, MN 55442-2920, Rx Only, Sterile EO. US Model Numbers: SC-014-135, SC-018-090, SC-035-065, SC-035-135, SC-014-150, SC-018-150, SC-035-090, and SC-035-150. Intended to guide and support a guide wire during access of the vasculature.

FDA Recall Z-0497-2010 · Posted 2009-12-23

Recalling FirmEv3, Inc.
Recall StatusTerminated
Product CodeDQY
Date Initiated2009-11-06
Date Terminated2011-01-05
Product Quantity70

Reason for Recall

Catheter may crack near the radiopaque marker band at the distal end of the device during use. Cracking of the catheter shaft in this location can result in material separation, and potential embolization.

Firm Action

Consignees were sent a Medical Device Recall letter, dated 11/6/09. The letter was addressed to Risk Manger or Cath Lab Manage. The letter provided the Issue Summary, identified the Affected Product, and stated the Required Action, which asked consignees to locate and remove the affected product. Questions should be directed to ev3 Customer Service at 800-716-6700.

Root Cause

Component design/selection

Distribution Pattern

Nationwide Distribution -- USA, including states of CA, FL, GA, IL, IN, IA, KS, KY, MI, MT, NJ, PA, SD, TN, TX, and WI.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →