Reusable, Self-Adhering TENS/NMES/FES electrodes Single Patient Use Only. An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.
FDA Recall Z-0499-2013 · Posted 2012-12-10
Advertisement
Recalling FirmCovidien LLC dba Uni-Patch
Recall StatusTerminated
Product CodeGXY
Date Initiated2012-02-09
Date Terminated2013-01-03
Product Quantity3594 pouches
Reason for Recall
Covidien is conducting a voluntary recall regarding specific lots of TENS electrodes because it was determined that during the manufacturing process the surface of the wires in these lots may have been exposed to an employee who had a minor cut.
Firm Action
Covidien sent An Urgent Product Recall letter, dated February 9, 2012, to all affected consignees. The letter identified the affected product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory and quarantine the affected product if found. The letter asked customers to fill in teh quantity remaining in their inventory and fax it to 651-565-5888. A Covidien Customer Service rep would contact the consignee for returning product. Consignees were also requested to notify their customers of the recall.
For questions regarding this recall call 508-261-8532.
Root Cause
Employee error
Distribution Pattern
Nationwide Distribution including CA, IL, MA, MN, and SD.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.