Dimension Vista Lactate Dehydrogenase (LDI) Flex reagent cartridge, an in vitro diagnostic test for the quantitative measurement of lactate dehydrogenase in human serum and plasma on the Dimension Vista System.Cat. No. K2054, Mat. No. 10464323
FDA Recall Z-0499-2019 · Posted None
Advertisement
Recalling FirmSiemens Healthcare Diagnostics, Inc.
Recall StatusTerminated
Product CodeCFJ
Date Initiated2018-10-11
Date Terminated2020-06-29
Product Quantity13435
Reason for Recall
Siemens Healthcare Diagnostics has confirmed that Dimension Vista Lactate Dehydrogenase
(LDI) may exhibit an erroneously elevated result when the preceding assay in the reaction
cuvette is Lactic Acid (LA). The magnitude of the elevation may be variable. Both quality control
and patient samples may be impacted. Incomplete removal of residual lactate dehydrogenase, a
component of the LA reagent, has been identified as the cause of the potentially elevated result. A falsely elevated result for lactate dehydrogenase may lead to additional investigations to
determine the etiology of tissue damage. Results would be correlated with clinical history and
presentation in addition to other diagnostic laboratory testing.
Firm Action
An Urgent Medical Device Correction were issued on October 11, 2018 to all Dimension Vista customers to notify them of the issue, the potential risk to health, and actions to be taken by the customer.
Customers were instructed to:
-Configure Dimension Vista System to eliminate the potential for cuvette carryover from lactic acid.
-If unable to separate the assays due to lack of additional server based on the the Dimension Vista model, run Lactate Dehydrogenase method in duplicates and contact Siemens if either result is discordant.
-Complete and return the Field Correction Effectiveness Check Form attached to the letter within 30 days.
-If you have received any complaints of illness or adverse events associated with the product listed in Table 1 of the letter, immediately contact your local Siemens Customer Care Center or their local Siemens Technical Support Representative
Root Cause
Process control
Distribution Pattern
Worldwide-domestic nationwide and foreign distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.