Recall StatusTerminated
Date Initiated2008-09-29
Date Terminated2011-06-02
Product Quantity720 kits
Reason for Recall
The recall was initiated after the firm investigation of customer complaints of falsely elevated results determined the product was not meeting product performance expectations.
Firm Action
The recall was initiated with Qualigen faxing a copy of the Product Notification letter, dated 9/29/2008, to all customers who have received any sublot of the affected product . Faxes were sent beginning 9/29/08. The recall letter informed the customers that the affected product was not meeting the firm's expectations for product performance. Customers were instructed to please stop using the affected lots immediately as they have been noted to produce falsely elevated results and they should confirm any result that appears to be inconsistent with "a patient's medical history, clinical examination, and other findings". In addition, they may wish to retest any patients that were run with a different FastPack lot.
Beginning on 9/30/08, all customers were contacted by phone to confirm that they received the recall letter.
The customers were shipped replacement products at no cost and as part of the product replacement process, they were instructed to return any unused Total PSA FastPacks/Kits of lot 0807002 to Qualigen.
Distribution Pattern
Worldwide Distribution --- including USA and countries of Canada, UK , Italy, Switzerland, Sweden, & South Korea.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source