Recall StatusTerminated
Date Initiated2009-10-27
Date Terminated2011-12-24
Product Quantity55,600
Reason for Recall
Medtronic is issuing a medical device correction for the Model 3550-39 Titan anchor due to the potential for lead migration as a result of insert separation within the anchor. The Medtronic Model 3550-39 Titan anchor accessory kit is designed for use with Medtronic percutaneous in-line
connector spinal cord stimulation leads. Medtronic has received reports of separation of the titanium
insert
Firm Action
Consignees were sent a Medtronic "Urgent: Medical Device Correction" letter on 10/27/09. The letter was addressed to "Dear Healthcare Professional" and was dated October 2009. The letter described the problem and the product as well as the Scope and Potential Severity and gave Recommendations. Enclosures with the letter included: Use By Date (UBD) location on the Titan anchor accessory kit product packaging, Titan anchor patient letter, Reply card and Return envelope. Questions are direct to Medtronic via email at
[email protected].
Root Cause
Nonconforming Material/Component
Distribution Pattern
All states in the United States including Puerto Rico.
Australia, New Zealand, Canada, Andurra, Belgium, Czech Republic, Denmark, Finland, France, Germany, Greece, Netherlands, Ireland, Israel, Italy, Kuwait, Lebanon, Luxembourg, Norway, Portugal, South Africa, Spain, Sweden, Switzerland, Tukey, United Kingdom.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.