RANDOX Total Bilirubin.
FDA Recall Z-0501-2024 · Posted 2023-12-06
Advertisement
Recalling FirmRandox Laboratories Ltd.
Recall StatusOpen, Classified
Product CodeJFM
Date Initiated2023-10-11
Product Quantity1644 units
Reason for Recall
Randox has had reports of elevated patient results using Plasma (lithium heparin) samples with Total Bilirubin reagent (TBil) Vanadate Oxidation Method.
Firm Action
Consignees were emailed an Urgent Device Correction letter on 10/11/2023. The notice instructed consignees to only use Total Bilirubin Vanadate Oxidation Method on Serum samples, to review results generated with Plasma samples in like with the clinical profiles of the patient, discuss the contents of the recall notice with your facility's Medical Director, and to return the provided response form completed to [email protected] within 5 working days. The recall notice is to be further distributed to all affected customers and those within consignee organizations, if further distributed. A second recall notification, dated 4/26/24, was sent to consignees identifying the reason for the product issue to be intralipid interference. Reagent IFUs have been updated with sample specific Plasma (lithium heparin) interference claims for intralipids, which can be found on www.randox.com. Consignees are asked to complete and return the provided response form to [email protected] within five working days.
Root Cause
Device Design
Distribution Pattern
US Nationwide distribution in the states of AR, GA, NC, TN, & PR.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement