Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

BIOFLO 8F SINGLE TITANIUM NON- FILLED; VALVED, UPN H965440250, Cat. No. 44-025 - Product Usage: Long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, chemotherapy, analgesics, nutritional therapy and blood products. The device is also indicated for blood specimen withdrawal.

FDA Recall Z-0502-2020 · Posted None

Recall StatusTerminated
Product CodeLJT
Date Initiated2019-09-18
Date Terminated2020-09-24
Product Quantity582

Reason for Recall

Snap lock connectors provided within implantable port kits may not meet dimensional specifications, and the catheter cannot be inserted into the snap lock connector prior to connection to the port chamber, resulting in a delay to the implantation procedure.

Firm Action

On September 18, 2019, AngioDynamics sent Urgent Voluntary Medical Device Letters to impacted consignees. Customers were asked to do the following: IMMEDIATELY o Stop using the product subject to recall. o Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location). o Segregate this product in a secure location for return to AngioDynamics, Inc. o Forward a copy of this recall notification to all sites to which you have distributed affected product. IN THE EVENT THAT THE PORT PRODUCT UNDER RECALL HAS BEEN UTILIZED (IMPLANTED) o The occurrence rate of an out of specification snap lock connector is less than 100%. o A port kit containing a properly sized snap lock connector allows the catheter to be securely attached to the port. o There are no actions required for an implanted port with a properly attached catheter. Customers were asked to complete and return the Reply Verification Form, and to package and return the recalled product.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution - US Nationwide in the state of CA, CT, FL, GA, IA, IN, KS, KY, MI, MO, MS, MT, NE, NJ, NY, OH, OK, OR, PA, TN, TX, WA, WI and country of Canada.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑