Aquapak Adaptor, 028 NEB, INTL, Product No. 403128; Aquapak Adaptor, 033 NEB, INTL, Product No. 403133; Aqaupak Nebulizer Adaptor 028, Sterile, Shelfpak, Product No. 031-28 & 031-28F, Aqaupak Nebulizer Adaptor 033, Sterile, Shelfpak, Product No. 031-33; Aqaupak Nebulizer Adaptor 033, Sterile, Japanese, Product No. 031-33J. Used in anesthesiology.

FDA Recall Z-0504-2015 · Posted 2014-12-12

Recalling FirmTeleflex Medical
Recall StatusTerminated
Product CodeCAF
Date Initiated2014-09-03
Date Terminated2017-08-14
Product Quantity3,193,548 ea

Reason for Recall

Possibility that the nebulizer adaptor packaging may have open or weak seals which may affect the sterility of the adaptor.

Firm Action

Customers were notified of the recall by letter on September 3, 2014. The letter requested that they discontinue distribution and quarantine the products and return the products to Teleflex. The letter also requested a sub-recall.

Root Cause

Packaging change control

Distribution Pattern

Worldwide Distribution-US (nationwide) including Puerto Rico and the states of AL, AK, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, WY, and the countries of United Arab Emirates, Australia, Belgium, Brazil, Canada, Chile, China, France, Great Britain, Japan, India, Indonesia, Israel, South Korea, Mexico, Philippines, Saudi Arabia, Singapore, Taiwan, Thailand, Uruguay, and Zambia.

Official Source