MC3 VitalFlow Console, REF 58100; Blood pump of ecmo
FDA Recall Z-0509-2026 · Posted 2025-11-14
Advertisement
Recalling FirmMedtronic Perfusion Systems
Recall StatusOpen, Classified
Product CodeQNR
Date Initiated2025-10-03
Product Quantity167 units
Reason for Recall
As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout.
Firm Action
Medtronic Cardiac Surgery issued an Urgent: Medical Device Correction notice to its consignees on 10/03/2025. The notice explained the issue, potential harms, and requested the consignee contact Medtronic Sales Representative to schedule the service event to their affected VitalFlow" Console. Please reference Technical Service Update (TSU) "MOD0209" when scheduling. The notice also indicated that the consignees use unaffected VitalFlow Console devices, if available, until the correction has been completed.
For questions regarding the communication, please contact your Medtronic Sales Representative.
Root Cause
Software in the Use Environment
Distribution Pattern
US Nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.