BrightView XCT Model 882482 Product Usage: BrightView XCT is a gamma camera for Single Photon Emission Computed Tomography (SPECT) and integrates with an attenuation device consisting of flat panel x-ray imaging components. BrightView XCT produces non-attenuation corrected SPECT images and attenuation corrected SPECT images with an x-ray transmission data that may also be used for scatter correction. The nuclear medicine images and BrightView XCT images may be registered and displayed in a fused format (overlaid in the same orientation) to provide anatomical localization of the nuclear medicine data. The BrightView XCT imaging system should only be used by trained healthcare professionals.

FDA Recall Z-0511-2018 · Posted 2018-02-03

Recall StatusTerminated
Product CodeKPS
Date Initiated2017-12-15
Date Terminated2020-04-10
Product Quantity1387 in total

Reason for Recall

During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning against the cart), or it may fall/slide all the way to the floor.

Firm Action

Philips sent an Urgent Field Safety Notice letter dated December 15, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local Philips representative.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide Distribution - US Nationwide

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.