BrightView Model 882480 Product Usage: BrightView and BrightView X are gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies. The cameras can be used to perform planar static, dynamic, gated, and total body studies, as well as circular-orbit and noncircular orbit SPECT studies, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors only).
FDA Recall Z-0512-2018 · Posted 2018-02-03
Advertisement
Recalling FirmPhilips Medical Systems (Cleveland) Inc
Recall StatusTerminated
Product CodeKPS
Date Initiated2017-12-15
Date Terminated2020-04-10
Product Quantity1387 in total
Reason for Recall
During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning against the cart), or it may fall/slide all the way to the floor.
Firm Action
Philips sent an Urgent Field Safety Notice letter dated December 15, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local Philips representative.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide Distribution - US Nationwide