Thermo Fisher Scientific/Microgenics brand CEDIA Tobramycin II Assay, Catalog Number 100018, Product is manufactured and distributed by Microgenics Corporation, located at Fremont, CA., part of Thermo Fisher Scientific The CEDIA Tobramycin II Assay is an in vitro diagnostic medical device intended for the quantitation of Tobramycin in human serum or plasma.

FDA Recall Z-0513-2014 · Posted 2013-12-16

Recalling FirmMicrogenics Corp
Recall StatusTerminated
Product CodeLFW
Date Initiated2013-10-16
Date Terminated2014-02-03
Product Quantity370 units

Reason for Recall

Microgenics Corp., part of Thermo Fisher Scientific, is recalling CEDIA Tobramycin II Assay (Lot Number: 60169764) due to calibration stability problem relating to two complaints received.

Firm Action

Microgenics Corp., part of Thermo Fisher Scientific sent an Urgent Medical Device Correction letter dated Octobere 16, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue use and discard. Customers with questions were instructed to contact [email protected]. For question regarding this recall call 1-800-232-3342.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide Distribution - USA including WI, UT, ND, MT, WA, CA, OK and Internationally to Australia, Canada, and Germany.