Immutest Cup CLIA Waived w/Adulteration, Item No. IMCA-7MT-W

FDA Recall Z-0513-2016 · Posted None

Recalling FirmAmeditech Inc
Recall StatusTerminated
Product CodeJXM
Date Initiated2015-11-16
Date Terminated2016-11-09
Product Quantity144 kits

Reason for Recall

Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

Firm Action

A Medical Device Labeling Correction letter was sent on 11/16/15 to distributors and end users to inform them that Ameditechs Drugs of Abuse Tests have shown reduced reactivity for up to two (2) of the nineteen (19) BZO compounds for which specificity information is provided in the product insert. The letter informs the customers that the particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide. Customers with questions are instructed to contact Ameditech at (844) 616-4453 or email at [email protected].

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide Distribution.