Terminated

Microlab AT, part numbers 135001, 135001OOC, 135002, 135003, 135005, 135006, 135006R, 135007, 135008, 135016, 135028, 135029, 936400, 936400BLU. Microlab AT+ and Microlab AT+2, part numbers 142000, 142001, 142001R, 142002, 142003, 142005, 142006, 142006R, 142007, 142008, 142009, 142015, 142017, 142035, 142036

FDA Recall Z-0514-06 · Posted 2006-02-14

Recalling FirmHamilton Co
Recall StatusTerminated
Product CodeJTC
Date Initiated2005-12-13
Date Terminated2008-05-07
Product Quantity1071 units

Reason for Recall

Heat sinking on power boards may not operate as intended and the board may cease to operate.

Firm Action

On 12/15/05, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the product be returned.

Root Cause

Other

Distribution Pattern

Product was released for distribution to 132 consignees in US, and 17 International consignees

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →