Babylog 8000 plus with software version 5.n. Neonatal Respirator-Continuous Ventilator

FDA Recall Z-0515-06 · Posted 2006-02-14

Recalling FirmDraeger Medical, Inc.
Recall StatusTerminated
Product CodeCBK
Date Initiated2005-12-28
Date Terminated2007-08-07
Product Quantity2261 units

Reason for Recall

Failure to generate an alarm when an unintended disconnection occurrs

Firm Action

Recalling firm has issued a medical device recall letter to consignees to the attention of the Director, Neonatal Intensive Care Unit via Certified Mail to alert them of additional monitoring requirements if a PEEP setting of less than 2.5 mbar is used

Root Cause

Other

Distribution Pattern

Product was shipped to medical facilities nationwide