Babylog 8000 plus with software version 5.n. Neonatal Respirator-Continuous Ventilator
FDA Recall Z-0515-06 · Posted 2006-02-14
Advertisement
Recalling FirmDraeger Medical, Inc.
Recall StatusTerminated
Product CodeCBK
Date Initiated2005-12-28
Date Terminated2007-08-07
Product Quantity2261 units
Reason for Recall
Failure to generate an alarm when an unintended disconnection occurrs
Firm Action
Recalling firm has issued a medical device recall letter to consignees to the attention of the Director, Neonatal Intensive Care Unit via Certified Mail to alert them of additional monitoring requirements if a PEEP setting of less than 2.5 mbar is used
Root Cause
Other
Distribution Pattern
Product was shipped to medical facilities nationwide