ADVIA Centaur Calibrator 30, 2 pack, 6 pack, and Ref 2 pack. Siemens Material Number (SMN): 10379810, 10379811, 10697756.
FDA Recall Z-0515-2018 · Posted None
Advertisement
Recalling FirmSiemens Healthcare Diagnostics, Inc
Recall StatusTerminated
Product CodeCHP
Date Initiated2017-06-07
Date Terminated2019-05-08
Product Quantity5992 units (US and OUS)
Reason for Recall
Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur Enhanced Estradiol (eE2) on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems when calibrating with Calibrator 30 kit lots ending in 21 (C3021) as compared to Calibrator 30 kit lots ending in 20 (C3020).
Firm Action
An Urgent Medical Device Recall letter dated June 7, 2017 was distributed to all customers in the United States and an
Urgent Field Safety Notice was distributed to all customers outside the United States who have
received ADVIA Centaur Systems Calibrator 30 kit lot ending in 21(C3021). These letters instruct
customers who operate the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT
systems to discontinue use of C3021 . The letters inform customers that Calibrator 30 kit lots
ending in 20 (C3020) and 23 (C3023) are not impacted and can be used. The letters inform
customers who operate ADVIA Centaur CP systems to continue using their inventory of C3021,
as they are not impacted by this issue.
All affected domestic customers were sent a hard copy of the Urgent Medical Device Recall via
Federal Express. Fax-back forms will be used to confirm that the customers received the Urgent
Medical Device Recall letter.
Root Cause
Under Investigation by firm
Distribution Pattern
US and OUS.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.