Terminated

LAERDAL Silicone Resuscitator (LSR), Preterm Size, containing an Intake/Reservoir Valve. Sold as Part Nos: 850050-Preterm Basic w/o Mask in Carton; 850051-Preterm Complete in Carton; 850053-Preterm Complete in Compact Case; 850055-Preterm Complete in Display Case; 855500-Preterm Update Kit. Responsible firm on the label: Distributed in the U.S. by Laerdal Medical Corporation, Wappingers Falls, NY, Made in Norway.

FDA Recall Z-0519-04 · Posted 2004-02-10

Recall StatusTerminated
Product CodeBTM
Date Initiated2003-12-23
Date Terminated2005-11-16
Product Quantity1,247 units

Reason for Recall

Potential for the Flap Valve component of the Intake/Reservoir Valve to become dislodged from its mount, rendering the resuscitator inoperative.

Firm Action

Letters issued on 1/14/2004. The recall involves the replacement of the Inner Part of the Intake/Reservoir Valve, which is part of the Laerdal Silicone Resuscitator. Recall is to the user level.

Root Cause

Other

Distribution Pattern

Product was shipped to approximately 264 customers nationwide, and 3 foreign consignees (Mexico, Brazil and Northern Mariana Islands). Product was also sold to a few exporters who shipped units to unknown countries. Customers include distributors and end users. There were 2,403 parts shipped to Laerdal Medical Canada, Ltd., Toronto, Ontario, Canada. That firm is responsible for conducting an independent recall, reportedly with the knowledge of the Canadian government.

Official Source

View FDA Recall →