Para-Pak Clean Vial-For the collection, transportation, preservation, and examination of stool specimens containing intestinal parasites Catalog Number: 900312
FDA Recall Z-0519-2024 · Posted 2023-12-08
Advertisement
Recalling FirmMeridian Bioscience Inc
Recall StatusOpen, Classified
Product CodeFMH
Date Initiated2023-10-19
Product Quantity571 Cases
Reason for Recall
Plastic vials contain a manufacturing defect may cause an increased risk of leakage and presents a risk of
incidental exposure to biological hazards associated with direct contact with the patient specimen.
Firm Action
Meridan issued Urgent Medical Device Correction letters via email on 10/19/23 to Distributors and End-Users. Letter states reason for recall, health risk and action to take:
Identify and segregate any remaining inventory.
REQUIRED ACTIONS FOR PARA-PAK CLEAN VIAL, CATALOG 900312, LOT 509940Q:
" Identify and segregate any remaining inventory.
" Discard remaining Inventory of Para-Pak Clean Vial, Catalog 900312, Lot 509940Q in accordance
with local, state, and/or federal regulations. Record quantity discarded.
o Contact Meridian Technical Support at 1-800-343-3858 to coordinate product replacement
activities.
" Complete and return the Customer Notification Form. This will indicate notification of this field
correction. Complete this form even if you have no remaining inventory.
REQUIRED ACTIONS FOR PARA-PAK Zn-PVA/10% FORMALIN, CATALOG 301012, LOT 509130Q:
" Identify and segregate any remaining inventory.
" For remaining inventory of Para-Pak Zn-PVA/10% Formalin, Catalog 301012, Lot 509130Q, inspect
each individual vial set for signs of leakage.
o Leakage would be able to be visually observed through inspection of the Ziplock-style bag containing the two vials of preservatives. Signs of leakage include identification of
preservative outside of the plastic vials or at the bottom of the Ziplock-style bag.
DO NOT remove vials that exhibit leaking from the Ziplock-style bag. Handle with care to
prevent direct contact of the preservative with the skin or eyes.
o Record the quantity of 2 vial sets that demonstrated signs of leakage.
" Discard the remaining inventory of Para-Pak Zn-PVA/10% Formalin, Catalog 301012, Lot 509130Q in accordance with local, state, and/or federal regulations. Record quantity discarded.
o Contact Meridian Technical Support 1-800-343-3858 to coordinate product replacement activities.
" Complete and return the Customer Notification Form. This will indicate notification of this field correction. Complete this form even if you have no remaining invento
Root Cause
Process control
Distribution Pattern
Nationwide Foreign: Italy
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.