Yukon OCT Spinal System Yukon Anti-Torque Alignment Tube - Product Usage: The Anti-Torque Alignment Tube (the Alignment Tube ) provides counter rotation to reduce the torque applied to the anatomy when the size 15 Torque limiting shaft (the shaft is used to tighten the Yukon Set Screw to the specified torque.
FDA Recall Z-0520-2021 · Posted None
Advertisement
Recalling FirmK2M, Inc
Recall StatusTerminated
Product CodeNKG
Date Initiated2020-11-03
Date Terminated2021-01-12
Product Quantity4 devices
Reason for Recall
Shafts do not pass through the inner cannula of the Anti-Torque Alignment Tube from Lot KTGG due to a manufacturing nonconformance. This could result in a delay of surgery or result in surgical intervention later on.
Firm Action
Our records indicate Stryker previously distributed catalog no. 7601-9001, Lot KTGG to your location in a Stryker Loaner Set. The nonconforming instrument was subsequently identified and quarantined during routine processing of loaner sets by Stryker at its Leesburg facility.
This communication is being made for notification purposes only. The only action requested of you is to complete the Business Reply form found on the following page and return it to the email addresses referenced in the form.
If you have any questions or concerns, please contact Stryker Regulatory Compliance at 201-749-8090. On behalf of Stryker we thank you sincerely for your help and support in completing the Business Reply Form and returning it to Stryker by November 6th, 2020.
Root Cause
Nonconforming Material/Component
Distribution Pattern
US Nationwide distribution.