Terminated

The product is a mobile 3-wheeled electric scooter, model HMV 220. The product contains either a red or blue base. The red base is identified with SKU #9153641905. The blue base is identified with SKU #9153641906. The product is distributed by the Invacare Corporation.

FDA Recall Z-0521-06 · Posted 2006-02-16

Recall StatusTerminated
Product CodeINI
Date Initiated2005-11-01
Date Terminated2012-05-08
Product Quantity8,213 total devices

Reason for Recall

The throttle spring, located in the hand throttle assembly, can fracture, resulting in the scooter not stopping when the user releases the throttle.

Firm Action

The recalling firm sent field correction instructions to their customers on or about 11/1/05.

Root Cause

Other

Distribution Pattern

The product was distributed to OHL/Invacare, Rancho Technology Park Bldg. B 8950, Rancho Cucamonga, CA 91730; OHL/Invacare, 1100 Whitaker Road, Plainfield, IN 46168; Invacare Corporation, 39400 Taylor Parkway North Ridgeville, OH 44039; and Invacare, Dallas Texas Distribution Center, 3460 South Watson Rd., Arlington, TX 76014.